99th ATF2 weekly meeting

November 26 14:30- 16:30, ATF LC meeting room, KEK.

International phone/webex meeting

Cross-checks of the large beta optics, Sha Bai (IHEP (China))

file , ppt, pdf (9 pages, 176KB)

Q : Beam size can be measured from um by IPBSM at Present. Is it OK?
A : 1um is fine, also it can be measured by carbon wire scanners.
Q : How much is the horizontal beam size?
A : It is about 10um.
C Okugi ; location of wire scanner and IPBSM are different. It makes 1.5 time at the wire scanner larger than IPBSM.
Q : Where does sextupole effect appear ?
A : It would become significant at beam size with 6 times of the nominal one, which is 200nm.
Q : How much is the energy spread assumed ?
A : It is 8 x 10^-4 .
Q : What program do you use for your analysis?
A : It is MAD.

Vibration measurement on the Shintake monitor etc., Benoit BOLZON (LAPP) , Takashi Yamanaka (The University of Tokyo)

file , ppt, pdf (17 pages, 1.5MB)

Q : Is the cooling water ON ?
A : It is OFF, but very small effect was seen.
Q : How much is the tolerance ?
A : tolerance < 10nm, which corresponds to 2% beam size error. The measured ones are 7 and 20nm at QD0 and QF1.
Q : When did you measure?
A : It was mid-night for clear transfer functions.
C : While upstream magnets have less tolerance than QD0 and QF1 for the nominal optics, how about is pushed beta optics ? The tight tolerance may extend to upstream more, say 10 m upstream.
Q : What is future measurement?
A : I will measure along the beam line.
Q : For GM -model tuning, whole set of measurement vertical and horizontally , both direction are needed ?
A : Yes, it is good especially for coupling in h and v . if time is limited, v is only fine too.