Participants: Terunuma, Tauchi, Okugi, Yamanaka, Oroku, Kamiya(WebEx)
Time: 17:00 ~ 18:00
Site: KEK ATF LC Container
Q: Is it possible to move a new Cherenkov detector remotely?
A: Yes, it is not impossible. However, since it is heavy,
it may be difficult with devices in hand.
Furthermore it is uncertain whether continue to use a Chenkov
detector.
Q: How long is the stroke of the collimator scanning stage?
A: 100 mm in horizontal and 50 mm in vertical. They are enough.
C: Cause of the laser position drift may be air temperature rather than
laser cooling water. When air temperature changes, the support of
the mirrors deforms and laser path changes.
C: Stabilization of the laser path will be needed to suppress the position
drift.